Sleep Technology (RPSGT) · EEG and PSG Instrumentation
Instrumentation, Amplifiers, Filters, Sampling, and Digital Signals
On this page 7 sections
In 30 seconds
Instrumentation is the chain that turns tiny physiological voltages into usable, storable, and reviewable signals. A signal is acquired by electrodes, amplified by a Differential amplifier Amplifies difference, rejects common signal Full entry → that rejects common-mode noise (measured by CMRR Common-mode rejection ratio Full entry →), shaped by high- and low-pass filters (plus a Notch filter Removes a narrow band (50/60 Hz) Full entry → for power-line interference), and digitized at a Sampling rate Samples per second Full entry → and Digital resolution Bits per amplitude sample Full entry → that must satisfy the Nyquist theorem Sample ≥ 2× highest frequency Full entry → to avoid Aliasing High frequencies appear as low Full entry →. Understanding Gain Amplification factor Full entry →, Sensitivity Voltage per unit display deflection Full entry →, impedance, Time constant Inverse measure of low-frequency filter Full entry →, and the difference between display settings and the underlying stored signal is essential to recognizing whether a tracing reflects physiology or instrument artifact.
Why this matters
Instrumentation errors masquerade as clinical findings, so understanding the signal chain is a data-quality and safety skill. A technologist who knows CMRR, filters, time constants, sampling rates, and aliasing can tell whether a deflection is brain activity, artifact, or a display-filter effect — and can verify settings meet technical specifications. Settings follow current AASM/manufacturer requirements and institutional protocol; verification is recognition, while interpretation remains with the physician. Accurate, well-verified recordings underlie every downstream decision.
The college version
1. Acquisition, impedance, and amplification
Signal acquisition Detecting the physiological voltage Full entry → is detecting the small physiological voltages at the electrode-skin interface and passing them to the amplifier. Electrode impedance Resistance at electrode-skin interface Full entry → is the resistance at that interface; if too high, the signal is degraded. Input impedance is the resistance the amplifier presents to the signal source; it must be high relative to electrode impedance so it does not load down and shrink the small signal. The differential amplifier amplifies the difference between two inputs while rejecting what they share. A common mode signal is one present at both inputs (such as 50/60 Hz interference); the common-mode rejection ratio (CMRR) quantifies how well the amplifier rejects it — higher is better. Gain is the amplification factor; sensitivity describes how much voltage is represented per unit of display deflection (e.g., µV/mm). In older analog systems, pen deflection described how far a pen moved per unit signal. Signal calibration applies a known input to verify the amplifier's response, so amplitudes are trustworthy.
2. AC vs. DC coupling and filters
An AC-coupled amplifier blocks slow or steady (DC) components and passes the faster changes typical of EEG, EOG, EMG, and ECG; a DC-coupled amplifier preserves slow or sustained changes and is used for respiratory effort, oximetry, or position, where the slow baseline carries meaning. Filters shape which frequencies pass. A high-frequency/low-pass filter attenuates frequencies above its setting (removing fast artifact); a low-frequency/high-pass filter attenuates frequencies below its setting (removing slow drift). The time constant is a related way of expressing the low-frequency filter — longer time constants pass slower activity. A notch filter removes a narrow band centered on the 50/60 Hz power-line frequency. Filters are necessary but can cause filter distortion — altering waveform shape or amplitude near the cutoff — so technologists must recognize when a filter, not physiology, is shaping the tracing.
3. Sampling, digital resolution, and aliasing
Modern systems digitize the analog signal. The sampling rate is the number of samples per second (Hz), and digital resolution is the number of bits representing each sample's amplitude (more bits = finer steps). The Nyquist theorem states the sampling rate must be at least twice the highest frequency of interest; otherwise aliasing occurs, making high-frequency activity appear falsely as a lower frequency. Technologists verify settings meet technical specifications (per current AASM and manufacturer requirements) and understand that display settings vs. underlying signal can differ: on-screen filters and gain determine what is seen during review, while the stored raw signal may contain more than the display shows.
How it works
- Physiological voltages are acquired at the electrode-skin interface.
- A differential amplifier amplifies the difference and rejects common-mode noise (CMRR).
- Coupling (AC vs. DC) and filters (high-pass, low-pass, notch) shape the frequency content.
- Calibration verifies gain and sensitivity with a known input.
- The signal is digitized at a rate and resolution satisfying Nyquist to avoid aliasing.
- Display filters and gain determine what is seen, distinct from the stored raw signal.
Common confusions
| Do not confuse | With | Difference |
|---|---|---|
| AC amplifier | DC amplifier | Blocks slow/DC vs. preserves slow/sustained |
| High-frequency/low-pass filter | Low-frequency/high-pass filter | Cuts fast vs. cuts slow |
| Gain | Sensitivity | Amplification factor vs. voltage-per-deflection |
| Input impedance | Electrode impedance | Amplifier-side vs. electrode-side resistance |
| Sampling rate | Digital resolution | Time (samples/sec) vs. amplitude (bits) |
| Time constant | Sampling rate | Frequency (low-pass) vs. time (digitization) |
| Display settings | Underlying signal | What is viewed vs. what is stored |
Memory aids
Signal chain: "A-C-F-S-D" — Acquire, amplify (CMRR rejects common mode), Filter, Sample (Nyquist), Display. Filters: "Low-pass kills the highs; high-pass kills the lows; notch kills the hum." Aliasing: "Sample too slow, highs look low."
Quick review
Topic Recap
- The signal chain is acquisition → differential amplification → filtering → sampling → display.
- Differential amplifiers reject common-mode noise (CMRR); input impedance must be high.
- AC coupling suits fast signals; DC coupling preserves slow baselines.
- Low-pass/high-pass/notch filters shape frequency content; time constant relates to the low-frequency filter.
- Sampling rate and resolution must satisfy Nyquist to prevent aliasing.
- Display settings differ from the underlying stored signal and must be understood during review.
Knowledge Check
- Why is a differential amplifier used for physiological signals?
- When is DC coupling used instead of AC coupling?
- What does the notch filter remove, and at what frequency?
- State the Nyquist theorem and what happens when it is violated.
- Why must a technologist distinguish display settings from the underlying stored signal?
Answers and Rationales
- Because it amplifies the difference between inputs while rejecting common-mode noise (CMRR). Why: Physiological signals are small, and interference arrives at both inputs.
- For slow or sustained signals — effort, oximetry, position — where the baseline carries meaning. Why: AC coupling would block that slow information.
- A narrow band centered on the 50/60 Hz power-line frequency. Why: It is the most common environmental interference.
- The sampling rate must be at least twice the highest frequency of interest; otherwise aliasing makes high frequencies appear lower. Why: Undersampling cannot faithfully represent fast activity.
- Because on-screen filters and gain determine what is viewed, while the stored raw signal may contain more. Why: Otherwise display settings could be mistaken for physiology.

Eli explains
The same idea, in plain words
Explain it like I’m 10
Recording a sleep signal is like recording a quiet conversation in a noisy room. The microphone (electrode) picks up both the voice and the noise; the amplifier is like a smart filter that boosts the voice while canceling the hum arriving at both ears equally; the "filters" are like tone controls cutting rumble and hiss; and "sampling" is like taking many quick snapshots per second so the sound is captured faithfully. Too few snapshots and fast sounds get misread as slow ones — that is aliasing.
The comparison stops being exact because these are precisely specified electrical and mathematical processes — differential amplification, common-mode rejection, filter corner frequencies, time constants, and bit depth — with defined limits and failure modes (distortion, aliasing) that must be verified against technical standards. A technologist who misunderstands these can mistake a filter artifact for brain activity.
Simple Example
60 Hz power-line hum contaminates every EEG channel. Because the hum is a common-mode signal present at both amplifier inputs, the differential amplifier cancels most of it (high CMRR), and a notch filter removes the rest — leaving a clean brain signal.
Worked example
- Acquire — the electrode-skin interface passes the voltage to the amplifier.
- Amplify — a differential amplifier boosts the difference and rejects common-mode noise (CMRR).
- Couple and filter — AC or DC coupling plus high-pass/low-pass/notch filters shape the signal.
- Calibrate — a known signal verifies gain/sensitivity.
- Digitize — the signal is sampled at a rate and resolution satisfying Nyquist.
- Display — on-screen filters and gain determine what is viewed, distinct from the stored signal.
This describes the conceptual signal chain, not step-by-step equipment configuration or settings.
Key takeaways
- High yield: Differential amplifiers reject common-mode noise; CMRR quantifies how well.
- High yield: AC coupling for EEG/EOG/EMG/ECG; DC coupling for effort, oximetry, position.
- High yield: High-frequency/low-pass removes fast artifact; low-frequency/high-pass removes slow drift.
- High yield: The notch filter targets 50/60 Hz power-line interference.
- High yield: Longer time constant = pass slower activity.
- High yield: Nyquist: sampling rate must be at least twice the highest frequency of interest.
- High yield: Aliasing makes high frequencies appear as low when undersampled.
- High yield: Display settings and the underlying stored signal are not the same thing.
Study tools & related lessonsYou’ll learn to · Key vocabulary · Related
You’ll learn to
- Describe the path of a physiological signal from acquisition to displayed waveform.
- Contrast AC and DC amplifiers and explain differential amplification and CMRR.
- Explain the roles of low-pass, high-pass, and notch filters, including time constant and filter distortion.
- Define sampling rate, digital resolution, the Nyquist theorem, and aliasing, and explain why display settings differ from the underlying recorded signal.
Key vocabulary
- Signal acquisition
- Detecting the physiological voltage
- AC amplifier
- Blocks slow/DC, passes fast changes
- DC amplifier
- Preserves slow/sustained changes
- Differential amplifier
- Amplifies difference, rejects common signal
- Common mode signal
- Signal present at both inputs
- CMRR
- Common-mode rejection ratio
- Input impedance
- Amplifier's resistance to the source
- Electrode impedance
- Resistance at electrode-skin interface
- Sensitivity
- Voltage per unit display deflection
- Gain
- Amplification factor
- Pen deflection
- Analog pen movement per unit signal
- Signal calibration
- Known input verifies response
- High-frequency/low-pass filter
- Attenuates fast activity
- Low-frequency/high-pass filter
- Attenuates slow activity
- Notch filter
- Removes a narrow band (50/60 Hz)
- 50/60 Hz interference
- Power-line contamination
- Time constant
- Inverse measure of low-frequency filter
- Sampling rate
- Samples per second
- Digital resolution
- Bits per amplitude sample
- Nyquist theorem
- Sample ≥ 2× highest frequency
- Aliasing
- High frequencies appear as low
- AASM technical-spec verification
- Confirming settings meet standards
- Filter distortion
- Filters altering waveform shape
- Display settings vs. underlying signal
- On-screen vs. stored raw signal
Educational content only. It is not medical, legal or professional advice. Found an error? Tell us.
